PRODUCTS : ELECTRONICS : MICRO 80
Available Standard Versions | |
---|---|
Micro 40 | |
Micro 40 M6/25 | Special top plate with M6 tapped holes on a 25 mm pitch |
Micro 60 | |
Micro 60 M6/25 | Special top plate with M6 tapped holes on a 25 mm pitch |
Micro 80 |
|
Micro 80 M6/25 |
Special top plate with M6 tapped holes on a 25 mm pitch |
Performance Specifications |
|
Isolation technology: | Halcyonics VarioControl technology based on piezoelectric type acceleration pickup, fast signal processing and electro-dynamic type force transducers. |
Control electronics: | Easy-to-navigate menu for all settings, second graphics display for vibration levels |
Force directions: | Active compensation in all six degrees of freedom |
Isolation performance: | > 5 Hz = 25 dB (94.4%); > 10 Hz = 40 dB (99.0%) |
Active bandwidth: | 0.6 – 200 Hz* |
Settling time: | 300 ms |
Max. correction forces: | Vertical ± 8 N; horizontal ± 4 N |
Load capacity: | 0 – 100 kg (0 – 200 lbs), Micro 80: 90 kg (195 lbs) |
Other Specifications |
|
Dimensions: |
See figure 2 |
Weight: | Micro 40: 27.8 kg (61.3 lbs) Micro 60: 43.3 kg (95.5 lbs) Micro 80: 55.0 kg (121.3 lbs) |
Table top material: | Micro 40: powder coated aluminium Micro 60: powder coated aluminium, honey-comb structure Micro 80: powder coated aluminium, honey-comb structure |
Top plate surface flatness: | Micro 40: ± 0.10 mm over complete surface Micro 60: ± 0.15 mm over complete surface Micro 80: ± 0.15 mm over complete surface |
Maximum compensation level: | 500 µm/s at 6 Hz and 60 kg (132 lbs)** |
Repeatability of load adjustment: | 120 µm |
Interface: | USB 1.1 standard |
Software: (for Microsoft Windows based PCs) |
• Activate/deactivate active vibration isolation by remote control • Monitor function – display sensor signals • Spectrum analyzer function – guidance to locate disturbing vibration sources • Location finder – guidance to find the ideal installation location in the lab |
Environmental and Operational Requirements | |
Electrical voltage: | 100 – 250 V / 47 – 63 Hz |
Power consumption: | 40 – 55 W |
Operating temperature: | 10 – 40°C (50 – 104 F) |
Relative humidity: |
0 – 60% |
Operating altitude: | < 2500 m (8100 ft) |
Certification | |
Electrical Safety: | CE certificated according to directive 89/336/EC |
EMC: | CE certificated according to directive 73/23/EEC |
* Floating table top is supported by steel springs; low-pass characteristics of spring-mass combination dominates the dynamic behaviour above 200 Hz. ** The maximum compensation level depends on several conditions, such as payload, frequency, load distribution and height of the payload. For that reason this value should be considered as an estimation. |